Passport V Operating Instructions 0070- 0-0704-02 5 - 5
Respiration Monitoring Respiration and CO
2
 Troubleshooting
5.4 Respiration and CO
2
 Troubleshooting
MESSAGE/PROBLEM REASON SOLUTION
Resp. Waveform Too Large Scales set inappropriately. Change lead selection.
Change Respiration scale.
Resp. Waveform Too Small Patient breathing is shallow 
or patient is turned on side.
Change lead selection.
Scale set inappropriately. Change respiration scale.
False Apnea Alarm Apnea delay may be 
improperly set.
Choose another apnea 
delay
Patient may be having 
frequent episodes of CVA.
Reposition electrodes to 
better detect respirations.
Scale size may be too low. Change Respiration scale
No Resp. Waveform or 
Rate Displayed
Respiration turned Off. 
(Off will be displayed in 
Resp. tile).
Turn respiration On 
Patient connected using 
ESIS choke cable.
Check that proper patient 
cable is used. Use approved 
non-ESIS patient cable.
Cable not connected.
CO
2
 not selected in 
Display Setup Menu.
Check cable.
Set CO
2
 in Display Setup 
Menu.
“CHK Lead” Message Increased impedance 
caused by one of the 
following:
Chest hair under 
electrodes.
Prep chest.
Dried electrode gel. Change electrodes.
Electrode off. Replace electrode.
Lead off. Replace lead.
Cracked lead wires.  Replace lead wires.
Poor skin prep. Clean and abrade skin 
before applying electrodes.
“CVA” Message Can be caused by shallow 
breathing or an apnea 
event.
Check the patient.
Patient HR and respiratory 
rate identical.
Adjust scales or leads if 
necessary.
0