Passport V Operating Instructions 0070- 0-0704-02 14 - 7
Messages Technical Alarm Messages
14.2.9 System Level
ALARM MESSAGE PRIORITY REASON SOLUTION
Low Battery 3 The battery capacity
is lower than the set
alarm limit value.
Change the battery, or use the
AC mains.
12V Too High 1 The monitoring of the
internal voltage finds
that 12V voltage is
too high.
Contact Technical Support.
12V Too Low 1 The monitoring of the
internal voltage finds
that 12V voltage is
too low.
Contact Technical Support.
5V Too High 1 The monitoring of the
internal voltage finds
that 5V voltage is too
high.
Contact Technical Support.
5V Too Low 1 The monitoring of the
internal voltage finds
that 5V voltage is too
low.
Contact Technical Support.
3.3V Too High 1 The monitoring of the
internal voltage finds
that 3.3V voltage is
too high.
Contact Technical Support.
3.3V Too Low 1 The monitoring of the
internal voltage finds
that 3.3V voltage is
too low.
Contact Technical Support.