Gas Module Troubleshooting Gas Monitoring (optional)
11 - 8 0070- 0-0704-02 Passport V Operating Instructions
10. If not already set, use the Display Setup Menu to select the gas waveforms to be
displayed.
11. If desired, the gas waveform speed and scale can be changed in the Gas Menu.
11.3.3 Gas Module 3 Pre-use Test
Prior to each use, perform the following test with the Gas Module 3 to verify that the gas
analyzer and sample system are functioning properly:
1. Verify that the appropriate water trap is properly installed and that the appropriate
sampling line is connected.
•DRYLINE
™
Adult/Pediatric water trap used with DRYLINE
™
Adult/Pediatric sampling
line (colorless Luer lock nut)
•DRYLINE
™
Neonatal water trap used with DRYLINE
™
Neonatal sampling line (blue
Luer lock nut)
2. Verify that the water trap container is less than half full.
3. Occlude the sampling line and verify that the occlusion alarm functions properly.
4. Breathe into the sampling line and verify that a CO
2
waveform is correctly displayed on
the monitor.
5. Sample room air for 30 seconds and verify that the monitor oxygen output is 21%
(± sensor inaccuracy).
11.4 Gas Module Troubleshooting
Refer to Chapter 14.0 Messages for Physiological Alarm Messages, Technical Alarm
Messages, and Prompt Messages.