Care and Disinfection of the Passport V Monitor User Maintenance
17 - 2 0070- 0-0704-02 Passport V Operating Instructions
17.2 Care and Disinfection of the Passport V Monitor
WARNING: Shut down the monitor and disconnect all power cords from
the outlet before cleaning.
The equipment should be cleaned regularly. Please consult your hospital’s policy for the
recommended frequency for cleaning and disinfecting equipment.
The exterior surfaces of the equipment may be cleaned with a clean and soft cloth, sponge or
cotton ball, dampened with either of the following cleaning solutions:
• Mild soap (diluted)
• Sodium hypochlorite bleach (10%)
• Isopropyl alcohol (70%)
• Super sani-cloth (0.5% quaternary ammonia + 55% isopropyl alcohol)
•Virkon
To avoid damage to the equipment:
• ALWAYS use solutions in accordance with the manufacturer's instructions.
• ALWAYS wipe off the excess cleaning solution with a dry cloth after cleaning.
• NEVER submerge the equipment into water or an cleaning solution, or pour or spray
water or any cleaning solution on the equipment.
• NEVER permit fluids to run into the casing, switches, connectors, or any ventilation
openings on the equipment.