14.3.5 NIBP Prompt Message
ALARM MESSAGE REASON SOLUTION
NIBP: Self Test Error Failed self-test.
Sensor or A/D sampling may
have an error.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: Communication Error NIBP module communicatuon
error, the command can not be
sent correctly.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: System Error System error. After start-up, the
inflating pump, A/D sampling
unit, pressure sensor have an
error or the pointer has an error
when the software is in running
process.
Power supply is not stable or
circuit failure leads to voltage
error.
Power cycle unit. If message
reappears, contact Technical
Support.
NIBP: Cuff Overpressure The hardware overpressure limit
has been exceeded.
Overpressure. The cuff pressure
exceeds 297 mmHg in ADU
mode, 240 mmHg in PED mode
or 147 mmHg in NEO mode.
Power cycle unit.
If message reappears, contact
Technical Support.
Pneumatic Leak Leakage. In Pneumatic check, air
leakage is found in hose.
Change the cuff.
NIBP: Unable to Measure Unable to make measurement
after three automatic retries.
Check Patient.
Retry measurement. If message
reappears, power cycle unit.
If message reappears, contact
Technical Support.
NIBP: Resetting... /
NIBP: Reset Failed Reset has failed. Power cycle unit.
If message reappears, contact
Technical Support.
NIBP: Idle Appears when NIBP
measurement is completed and
enters the wait status.
This is normal operation.
Press 'Start' key to start a
measurement, the message is
cleared.
NIBP: Press the START key Appears when NIBP Interval is
set to anything other than Off or
STAT.
This is normal operation.
Press 'START' key to start
automatic measurement.
NIBP: Manual Mode Displayed while system is idle.
Note: This is not displayed while
in the interval mode.
Press START to take a single
measurement. Select an
interval and start timed
measurements.
NIBP: Interval Displayed during the interval
between two timed
measurements.
Press STOP to suspend timed
measurements.
Change timer to OFF to stop
timer.
NIBP: Measuring Displayed during a
measurement.
Cuff pressure is also displayed.
Press STOP to suspend a
measurement and deflate the
cuff.