Performance Characteristics Appendix
19 - 14 0070- 0-0704-02 Passport V Operating Instructions
• Refresh Rate: 1s
• Filter: Optional three bandwidths:
• -3 dB point at 8 +/- 1 Hz,
• -3 dB point at 12 +/- 1 Hz,
• -3 dB point at 20 +/- 1 Hz,
• Alarm: Systolic pressure, diastolic pressure, mean pressure.
Automatic Scaling: Supported
• Data Storage: Physiological parameter: Systolic pressure, Diastolic pressure, Mean
pressure
19.8.4 NIBP Performance and Functional Characteristics
The NIBP function is capable of providing systolic, diastolic, and mean blood pressure
measurements in Neonate, Pediatric, and Adult modes non-invasively, using a blood pressure
cuff.
The NIBP function is in accordance with the requirements of EN 60601-2-30, EN 1060-1,
EN 1060-3, and SP 10.
Systolic Pressure Measurement
Diastolic Pressure Measurement
Static Pressure Measurement
Accuracy: Mean error is less than ± 5 mmHg, Standard
Deviation is less than ± 8 mmHg.
Range:
Adult Mode 55 to 235 mmHg
Pediatric Mode 55 to 160 mmHg
Neonatal Mode 45 to 120 mmHg
Accuracy: Mean error is less than ± 5 mmHg, Standard
deviation is less than ± 8 mmHg.
Range:
Adult Mode 30 to 200 mmHg
Pediatric Mode 30 to 150 mmHg
Neonatal Mode 20 to 100 mmHg
Range: 0-300 mmHg
Static Accuracy: ± 3 mmHg
0