NIBP Menu NIBP Monitoring
7 - 6 0070- 0-0704-02 Passport V Operating Instructions
Interval Off (default)
STAT, 1 min, 2 min,
3 min, 5 min, 10 min,
15 min, 20 min,
30 min, 1 hr, 2 hr, 4 hr
Select to change NIBP measurement interval.
Off: manual mode. The message "NIBP: Manual
Mode" is displayed at the bottom of the screen.
STAT: STAT measurement. The message "NIBP:
STAT" is displayed at the bottom of the screen.
Note: The Interval cannot be changed during an
NIBP measurement or the five minutes STAT
measurement.
NIBP End Tone On (default)
Off
Select to turn NIBP End Tone On or Off.
On: A dual beep tone sounds on completion of a
successful NIBP measurement.
Off: Tone does not sound after the measurement.
Note: A single beep tone will always sound in the
event of an unsuccessful measurement, after all
retries have been attempted.
The NIBP End Tone does not sound if the user stops
the measurement by pressing the NIBP STOP key.
NIBP End Tone audio volume is controlled by the
alarm audio volume.
NIBP Menu
MENU ITEM SELECTIONS COMMENTS
0

Table of Contents