Passport V Operating Instructions 0070- 0-0704-02 19 - 37
Appendix Electromagnetic Capability
TABLE 19-3
GUIDANCE AND DECLARATION - ELECTROMAGNETIC IMMUNITY
The Passport V is intended for use in the electromagnetic environment specified below. The customer or
the user of the Passport V should assure that it is used in such an environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
Portable and mobile RF communications
equipment should be used no closer to
any part of the Passport V, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation
distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms d = 1.2 x
Radia
ted RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m d = 1.2 x 80 MHz to 800 MHz
d = 2.3 x 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
a
should be less than the
comp
liance level in each frequency
range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE: At 80 MHz and 800 MHz, the higher frequency range
applies.
NOTE: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
NOTE: The device that intentionally receives RF electromagnetic
energy at the exclusion band (2409.5MHz-2464.5MHz) is
exempt from the wireless performance requirements, but
remains safe.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theo-
retically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electro-
magnetic site survey should be considered. If the measured field strength in the location in which the
Passport V is used exceeds the applicable RF compliance level above, the Passport V should be observed to ver-
ify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Passport V.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
0