Passport V Operating Instructions 0070- 0-0704-02 1 - 5
General Product Description Keys and Front Panel
7. Zero All (with IBP option)
Press this key to set the current pressure for all invasive pressure channels to zero. This key
does not affect any channels monitoring pressure. If the zero process is not successful, one of
the following messages is displayed: Pulsatile Pressure. Cannot Zero! or Pressure
Overrange. Cannot Zero!
8. Limits
Press this key to display the Alarm Settings Menu. The Alarm Settings Menu provides
access to view or change alarm values.
9. Silence All
Press this key to suspend audio alarms on all parameters. The alarms remain suspended for a
user selected amount of time. This amount of time is set in the Alarm Setup Menu. While
the alarms are suspended, an Alarm Silence icon and message All Alarms Silenced For
X:XX mins display in the upper area of the screen. X:XX is the time remaining in minutes
and seconds. Press this key again during the suspended alarm time to re-enable the audio
alarm. If the suspend time was set to Permanent in the Alarm Setup Menu, the message
All Alarms Silenced Permanently is displayed. Note that the Permanent selection
will be available only if Enable Silence All Permanent Selection is set to “Yes” in the
Installation Menu.
10. Silence
Press this key to suspend audio alarms on all currently alarming parameters. The alarms
remain suspended for a user selected amount of time as set in the Alarm Setup Menu or
until the alarm condition is no longer present. Any new alarms that occur while the alarm
tone is silenced will disable the silence and sound the alarm tone. While the alarms are
suspended, an Alarm Silence icon and message Alarm Silenced For X:XX mins display
in the upper area of the screen.
11. Strip
Press this key to initiate a printout to the selected device.
• If the print destination is the local printer, then pressing this key will produce a print strip
of up to three (3) waveforms. The print length (16 or 32-seconds) can be set in the Print
Setup menu. Pressing this key during a print job will abort the strip printout.
• If the print destination is a remote Central Station, then pressing this key will initiate a
printout at the Central Station.
• If the print destination is a remote printer, then pressing this key will initiate a printout at
the remote printer.
12. Cont ECG
Press this key to initiate a continuous ECG 1, 2, or 3 waveform printout from the internal
printer. Press this key again to abort printing. The first three ECG waveforms displayed on the
screen are printed continuously in real-time. If only one or two ECG waveforms are displayed
on the screen, they will be printed.
0