Passport V Operating Instructions 0070- 0-0704-02 3 - 7
Patient Management Remote View
All Remote View controls are contained within the Remote View Menu. Opening other
menus from the monitor’s keypad, or pressing the Normal Screen key, will close the
Remote View Menu.
The Remote View Menu displays data as follows:
The waveforms that are displayed are user-selectable through the Remote View
Setup Menu, as described in the table on page 3-8. Corresponding numeric tiles
are displayed to the right of the waveforms. The numeric data for a maximum of fifteen
(15) parameters that are not related to the waveforms is displayed in the “Numeric
Data Area” located directly below the waveforms.
The colors of the waveforms and all numeric data correspond to the settings of the host
monitor (not the remote monitor).
When numeric data for a particular parameter is unavailable, dashes (– – –) are
displayed.
If a high or low alarm limit has not been set, an Alarm Off icon is displayed.
Remote View alarms follow the alarm settings at the remote monitor, displaying in a
priority-appropriate, reverse-video color.
If the remote monitor is placed into Standby mode, the message “In Standby” is
displayed in the Remote View Window.
NOTE: Respiration alarm text messages are also displayed in the
displayed in the message area (upper part) of the Remote
View Menu window as follows:
The Remote View Menu also provides the following menu choices: Normal Screen,
Select Beds, Setup, Remote Silence, Remote Silence All, and Remote Print.
1. Normal Screen — this selection removes the Remote View Menu from the display.
2. Select Beds — this selection opens the Select Beds menu from which beds may be
selected to be included in the care group. All networked beds will be in que but only 8
can be selected.
If Apnea is detected, the message “APNEA” is displayed.
If CVA is detected, the message “CVA” is displayed.
If high impedance is detected from the ECG electrodes, the
message “CHK LEAD” is displayed.
0
NOTE: Remote View function is not available between wireless
monitors, nor for wireless monitors to view wired monitors.
Remote View function is available between wired monitors,
or for wired monitors to view wireless monitors.