Menus: ECG Main and Submenus ECG Monitoring
4 - 14 0070- 0-0704-02 Passport V Operating Instructions
WARNING: Ensure that the conductive parts of ECG electrodes do not
contact other conductive parts, including earth ground.
CAUTION: To avoid possible damage to the Passport V, and to provide
protection against the effect of the discharge of a cardiac
defibrillator and against burns, use only approved ECG
cables and approved accessories listed in the Accessories
chapter.
CAUTION: Line Isolation Monitor transients may resemble actual
cardiac waveforms, thus inhibiting heart rate alarms. Check
lead wires for damage and ensure good skin contact prior
to and during use. Always use fresh electrodes and follow
proper skin preparation techniques.
WARNING: This device is not intended for direct cardiac application.
ECG Cable Auto (Default)
3 Lead
5 Lead
Select to change the ECG cable type being used.
ECG Noise Audio On (Default)
Off
Select to turn ECG Noise audio signal on or off.
On: An audio alarm tone is triggered when ECG
Noise is detected.
ECG Noise Delay 3 to 30 seconds
Default = 5 seconds
Select to change the number of seconds to delay the
ECG Noise Alarm.
ECG Setup Menu
MENU ITEM SELECTIONS COMMENTS