Analog/Digital Converter (ADC) Conguration
Condition The weighing platform must already be connected.
Open the Menu Access Switch The menu access switch is located on the back of the indicator right next to the
weighing platform connection.
t Remove the cap.
t Slide the switch to the left (= “open“ position).
Activating the Service Mode t See “Service Mode“ on page 17.
Conguration t Select weighing platform “WP 1.“
t If the Internal“ setting is not already activated (marked by “o“), select
the setting using the “Q“ or “q“ soft key and conrm with “O.“
y The message Function active“ appears briey. The “WP 1 internal“
menu will then open.
t Open the ADC Configuration menu.
t Select the desired conguration data record using the “Q“ or “q“ soft key:
Standard“ or “Verifiable“ (veriable conguration).
The default setting depends on the data record.
t Open the menu for conguring A/D converter parameters. In this example, the
menu for ADC conguration of a weighing platform for use in legal metrology
is opened.
If Standard“ was selected previously, then the “Accuracy tass
is not displayed.
t Open the rst menu item.
For a standard conguration, the “Ranges“ menu item, for a veriable
conguration, the “Accuracy class“ menu item.
When the “Veriable” conguration is active, always select the “Accuracy
class
“ menu item rst.
t Set accuracy class l/m. The “o“ symbol should mark the setting, if required
conrm using the “l“ soft key.
t Press the “o“ soft key to exit the menu item.
t Open the “Ranges“ menu item.
Operating Instructions Combics Complete Scales 27
Configuring Weighing Platforms