Analog/Digital Converter (ADC) Conguration
Condition The weighing platform must already be connected.
Open the Menu Access Switch The menu access switch is located on the back of the indicator right next to the
weighing platform connection.
t Remove the cap.
t Slide the switch to the left (= “open“ position).
Activating the Service Mode t See “Service Mode“ on page 17.
Conguration t Select weighing platform “WP 1.“
t If the “Internal“ setting is not already activated (marked by “o“), select
the setting using the “Q“ or “q“ soft key and conrm with “O.“
y The message “Function active“ appears briey. The “WP 1 internal“
menu will then open.
t Open the ADC Configuration menu.
t Select the desired conguration data record using the “Q“ or “q“ soft key:
“Standard“ or “Verifiable“ (veriable conguration).
The default setting depends on the data record.
t Open the menu for conguring A/D converter parameters. In this example, the
menu for ADC conguration of a weighing platform for use in legal metrology
is opened.
If “Standard“ was selected previously, then the “Accuracy tass“
is not displayed.
t Open the rst menu item.
For a standard conguration, the “Ranges“ menu item, for a veriable
conguration, the “Accuracy class“ menu item.
When the “Veriable” conguration is active, always select the “Accuracy
class
“ menu item rst.
t Set accuracy class l/m. The “o“ symbol should mark the setting, if required
conrm using the “l“ soft key.
t Press the “o“ soft key to exit the menu item.
t Open the “Ranges“ menu item.
Operating Instructions Combics Complete Scales 27
Configuring Weighing Platforms