In the example shown here, “Single range mode“ has been selected
(marked by “o“).
t To select a dierent weighing range conguration, use the “Q“ or “q“ soft key
to select the corresponding line and open the selected menu using the “O“ soft
key.
y The selected weighing range conguration is now active. When you return
from the input menu for entering the weighing range parameters, the new
range conguration is marked by a “o.“
For more information about range conguration, please see “Setting Parameters for
ADC Conguration.“
The default values displayed depend on the data record loaded and might have
to be changed. In the example shown here, the A/D conguration is set with a
“Veriable” data record in single-range mode.
Single-range Mode t Select the individual input elds using the “
Q“ or “q“ soft key.
t For numeric input: use the 0 ... 9 keys and the . key (decimal point).
Make corrections using c.
t Conrm using the “l“ soft key.
If other parameters follow the one just entered, the highlight bar is
automatically positioned on the next input eld.
To cancel numbers entered: press the “ESC“ soft key.
t Use the “
o“ soft key to go to the next menu level.
y This will apply all parameters.
t Press M or “oo“ to exit the Setup menu.
0 . 0 0 2
In the example shown here, a single-range scale in “Veriable” conguration
with a maximum capacity of 6.000 kg is modied; the verication scale interval
e is changed from 0.001 kg to 0.002 kg, in accordance with the maximum
permitted value of 3000 verication scale intervals. Press the “l“ soft key to
conrm the changed value.
The highlight bar is automatically positioned on the “Min. cap.“ eld.
The following values apply for the minimum load for veriable scales:
For class l: Min. cap. = 20 e
For class m: Min. cap. = 10 e
A verication scale interval that is changed, therefore, also aects the minimum load.
Changing the verication scale interval “e“ is automatically applied to the “Min. cap.“
You can also change this value manually:
In the example, the minimum capacity must be increased to 0.04 kg for class l.
t Press the following keys in sequence: 0 . 0 4 0 and conrm using
the “l.“
 y The highlight bar is automatically positioned on the “Max. cap.“ eld.
t The value for the maximum capacity is not changed. For this example, the
input of parameters for single-range mode in the “Veriable” conguration is
now concluded.
t Use the “o“ soft key to go to the next menu level.
28 Operating Instructions Combics Complete Scales