In the example shown here, “Single range mode“ has been selected
(marked by “o“).
t To select a dierent weighing range conguration, use the “Q“ or “q“ soft key
to select the corresponding line and open the selected menu using the “O“ soft
key.
y The selected weighing range conguration is now active. When you return
from the input menu for entering the weighing range parameters, the new
range conguration is marked by a “o.“
For more information about range conguration, please see “Setting Parameters for
ADC Conguration.“
The default values displayed depend on the data record loaded and might have
to be changed. In the example shown here, the A/D conguration is set with a
“Veriable” data record in single-range mode.
Single-range Mode t Select the individual input elds using the “
Q“ or “q“ soft key.
t For numeric input: use the 0 ... 9 keys and the . key (decimal point).
Make corrections using c.
t Conrm using the “l“ soft key.
If other parameters follow the one just entered, the highlight bar is
automatically positioned on the next input eld.
To cancel numbers entered: press the “ESC“ soft key.
t Use the “
o“ soft key to go to the next menu level.
y This will apply all parameters.
t Press M or “oo“ to exit the Setup menu.
0 . 0 0 2
In the example shown here, a single-range scale in “Veriable” conguration
with a maximum capacity of 6.000 kg is modied; the verication scale interval
e is changed from 0.001 kg to 0.002 kg, in accordance with the maximum
permitted value of 3000 verication scale intervals. Press the “l“ soft key to
conrm the changed value.
The highlight bar is automatically positioned on the “Min. cap.“ eld.
The following values apply for the minimum load for veriable scales:
– For class l: Min. cap. = 20 e
– For class m: Min. cap. = 10 e
A verication scale interval that is changed, therefore, also aects the minimum load.
Changing the verication scale interval “e“ is automatically applied to the “Min. cap.“
You can also change this value manually:
In the example, the minimum capacity must be increased to 0.04 kg for class l.
t Press the following keys in sequence: 0 . 0 4 0 and conrm using
the “l.“
y The highlight bar is automatically positioned on the “Max. cap.“ eld.
t The value for the maximum capacity is not changed. For this example, the
input of parameters for single-range mode in the “Veriable” conguration is
now concluded.
t Use the “o“ soft key to go to the next menu level.
28 Operating Instructions Combics Complete Scales
Configuring Weighing Platforms