Multi-interval Mode The illustration shows an example of the input menu opened for multi-interval
mode range conguration. This example shows the parameters for a scale in
“Veriable” conguration, with 2 weighing ranges and a maximum capacity of
6.000 kg.
– Range 1: 0…3.000 kg with e1 = 0.001 kg
– Range 2: 3.002…6.000 kg with e2 = 0.002 kg
t Enter the verication scale interval for Range 1 in the “
E“ input eld.
The minimum capacity for a class l scale must be set to 0.02 kg.
t Use the “o“ soft key to go to the next menu level.
y The active range conguration is marked with “o.“
t Use the “o“ soft key to go back to the “Verifiable“ menu.
Selecting Units t Use the “q“ and “O“ soft key to open the “Available units“
menu item.
This menu lets you enable or disable the weight units available under “Weight
unit 1,“ “Weight unit 2“ and “Weight unit 3.“
t Select the respective unit using the “Q“ or “q“ soft key and conrm using the
“l“ soft key.
In most cases, you will not need to change dened values.
Available weight units are marked by a *.
The weight unit used for conguration of weighing ranges cannot be blocked.
t Use the “o“ soft key to go back to the “Internal“ menu.
t Use the “q“ and “O“ soft key to open the “Calibration/
adjustment
“ menu.
t Use the “q“ and “O“ soft key to open the “Calibration/
adjustment unit
“ menu item to dene the weight unit for
calibration and adjustment. In most cases, you will not need to change dened
values .
y All units are displayed in the menu that are activated in “Available
units
.“ The current setting is marked by a “o.“
Operating Instructions Combics Complete Scales 29
Configuring Weighing Platforms