t To change the calibration/adjustment unit, select the unit using the “Q“ or “q
soft key and conrm using the “l“ soft key.
t Use the “o“ soft key to go back to the “Internal“ menu.
t Use the “Q“ soft key to select the “ADC configuration“ menu item.
t Use the “O“ and “q“ soft key to select the “Save configuration
data
“ menu item.
t To save the conguration, use the “Q“ soft key to select “Yes“ and conrm
using the “l“ soft key.
y The message “Function activated“ appears briey.
The program then returns automatically to the regular weighing mode.
To not save the conguration:
t Press the “o“ soft key to exit the menu.
y The program returns to the next higher menu level.
To not save data:
t Press the “o“ soft key. The program returns to the WP1:Internal.
t Slide the menu access switch to the right (= “closed“ position) and reattach the
cap.
y The device is now in normal weighing mode.
h
Once ADC conguration has been completed, an adjustment of the weighing
platform (calibration/adjustment and linearization) must be carried out (see
“Calibration/Adjustment without Weights“ and “External Linearization“).
30 Operating Instructions Combics Complete Scales
Configuring Weighing Platforms