t Press e to turn o the device.
t Press e turn the device back on.
y The Minebea Intec logo is displayed briey, after which the device is in normal
weighing mode.
t Press ( to unload and zero the scale.
t Press J to start the external adjustment.
y The display “C.EXT.U“ appears briey.
Note
If the display of geographical data (elevation and latitude or gravity) is activated,
this data is displayed and conrmed each with a press of the J key (you can
cancel the calibration/adjustment process using the ( key). See also “Entering
Geographical Data for Use in Legal Metrology“ in this chapter.
Procedure t Press J to conrm the displayed value or press ( to cancel
the adjustment.
y The target value of the required adjustment weight (6.000 kg in the example) is
displayed as a negative value.
t Place the required adjustment weight on the platform.
If the calibration/adjustment sequence is set to “Cal. then auto adjust”, refer to
the note under “External Calibration/Adjustment with Factory-Dened Weight
(Default Weight).”
y After a brief pause, the dierence since the last span adjustment is displayed
(calibration).
t If you only want to perform a calibration, press the ( key to cancel the
calibration/adjustment procedure.
t Press J to adjust the scale.
y At the conclusion of the calibration procedure, the calibration weight is dis-
played as a positive value.
When calibration/adjustment is complete, the GMP-compliant printout shown on
the left is generated. If the adjustment procedure is canceled (only calibration is
performed), the last two lines, “External calibration” and “Di. + 0.000 kg” are not
printed.
t Unload the scale.
t Press e to turn o the device.
t Press e turn the device back on.
y The Minebea Intec logo is displayed briey, after which the device is in normal
weighing mode.
--------------------
24.02.2016 10:15
Typ CAW3S
Ser.no. 12345678
Vers. H0 111.031115
BVers. 01-63-02
Ser.no. A 12345678
--------------------
Ext. Calibration
Nom + 6000 g
Diff. + 10 g
Ext. adjustment
Diff. + 0 g
--------------------
24.02.2016 10:15
Name:
--------------------
42 Operating Instructions Combics Complete Scales
Configuring Weighing Platforms