EasyManua.ls Logo

Minebea Intec CAS-SPC Series - User-Defined External Adjustment; External Adjustment Procedure

Minebea Intec CAS-SPC Series
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
t Press e to turn o the device.
t Press e turn the device back on.
y The Minebea Intec logo is displayed briey, after which the device is in normal
weighing mode.
t Press ( to unload and zero the scale.
t Press J to start the external adjustment.
C.EXT.U
C.EXT.USE.
y The display “C.EXT.U appears briey.
Note
If the display of geographical data (elevation and latitude or gravity) is activated,
this data is displayed and conrmed each with a press of the J key (you can
cancel the calibration/adjustment process using the ( key). See also “Entering
Geographical Data for Use in Legal Metrology“ in this chapter.
Procedure t Press J to conrm the displayed value or press ( to cancel
the adjustment.
y The target value of the required adjustment weight (6.000 kg in the example) is
displayed as a negative value.
t Place the required adjustment weight on the platform.
If the calibration/adjustment sequence is set to “Cal. then auto adjust”, refer to
the note under “External Calibration/Adjustment with Factory-Dened Weight
(Default Weight).”
y After a brief pause, the dierence since the last span adjustment is displayed
(calibration).
t If you only want to perform a calibration, press the ( key to cancel the
calibration/adjustment procedure.
t Press J to adjust the scale.
y At the conclusion of the calibration procedure, the calibration weight is dis-
played as a positive value.
When calibration/adjustment is complete, the GMP-compliant printout shown on
the left is generated. If the adjustment procedure is canceled (only calibration is
performed), the last two lines, “External calibration” and “Di. + 0.000 kg” are not
printed.
t Unload the scale.
t Press e to turn o the device.
t Press e turn the device back on.
y The Minebea Intec logo is displayed briey, after which the device is in normal
weighing mode.
--------------------
24.02.2016 10:15
Typ CAW3S
Ser.no. 12345678
Vers. H0 111.031115
BVers. 01-63-02
Ser.no. A 12345678
--------------------
Ext. Calibration
Nom + 6000 g
Diff. + 10 g
Ext. adjustment
Diff. + 0 g
--------------------
24.02.2016 10:15
Name:
--------------------
42 Operating Instructions Combics Complete Scales
Configuring Weighing Platforms