Procedure t Remove the cap.
t Slide the menu access switch to the left (= “open“ position).
t Activate the Service mode, see “Service Mode.“
t Select weighing platform “WP 1“ in the “Device Parameters“
menu item.
t If the “Internal“ setting is not already activated (marked by “o“), select
the setting using the “Q“ or “q“ soft key and conrm with “O.“
y The menu for the “WP-1 INTERNAL“ device parameters is displayed.
t Use the “Q“ or “q“ soft key to select and open the “Calibration/
adjustment
“ menu using “O.“
t Use the “Q“ or “q“ soft key to select and use “O“ to open the “Adjust
without weights
“ menu.
t Use the “O“ soft key to open the “Input parameters“ menu.
y The Input menu is displayed.
Enter the nominal capacity and resolution of the load cells in kg and the
sensitivity of the load cells in mV/V in the corresponding input elds.
The maximum capacity is usually less than the value to be entered in the
“Adjust without weights“ menu for the nominal capacity of
the load cell, as load cells carry additional weight (e.g., a weighing platform).
44 Operating Instructions Combics Complete Scales
Configuring Weighing Platforms