90
Setting value is good.
Is the intended
operation proceeded?
Contact output chatters.
Contact output isnt ON.
Connecting wires is
correct?
Please connect the wires surely according
to the paragraph 13-4.
Please confirm and change the setting
value according to the chapter 11.
Comparator operation check
Please confirm and change the
comparative operation of comparator, and
the target of comparison operation
according to the chapter 11.
Apply the setting for hysteresis according
to the chapter 11.
Following causes can be considered:
Failure in contact by using out of rated
capacity of contact.
Failure in customers input circuit.
Please confirm the contact output
according to the paragraph 10-4.
Inform Minebea about the contents of
failure and situation at site in details.
YES
YES
YES
NO
NO
NO
NO
YES
YES
NO