Check of BCD output
Connecting wires is correct?
NO
YES
NO
YES
NO
YES
NO
YES
The BCD output does
not change.
The intended value is
not output.
Output except BCD
Please connect surely with the
external instrument according to
the paragraph 13-2.
Please open the HOLD input.
Please select the BCD output target
according to the paragraph 13-2.
Please match setting concerning to
the BCD output to the specification
of an external connected equipment
according to the paragraph 13-2.
Inform Minebea about the contents of
failure and situation at site in details.