95
Check of RS-232C interface
The value intended
in the stream mode is not
transmitted.
NO
YES
Please select the output target at
the stream mode.
Connecting wires is correct?
NO
YES
Please use correct command/data
referring to paragraph 13-3.
The communication protocol
is set correctly.
NO
YES
Please make host's flow control
none.
Command/data is set
correctly.
NO
YES
Please match the setting of
function correctly to the
specification of the host according
to the paragraph 13-3.
The flow control is specified
by the host.
NO
YES
Please connect the wires surely
according to the paragraph 13-3.
Inform Minebea about the contents of
failure and situation at site in details.