96
Check of RS-422 interface
NO
Connection wires is correct?
YES
Please connect wires surely with an
external equipment according to the
paragraph 13-3.
NO
The address number is set
correctly.
YES
Please change RS-422/485
correctly.
NO
The communication
protocol is set correctly.
YES
Please make host's flow control
none.
NO
Command/data is set
correctly.
YES
Please set the address number
correctly referring to the paragraph
13-3.
NO
The flow control is specified
by the host.
YES
Please match the setting of
function correctly to the
specification of the host.
NO
The change of
RS-422/485 is set
correctly.
YES
Please use correct command/data
referring to the paragraph 13-3.
Inform Minebea about the contents of
failure and situation at site in details.