Check of RS-485 interface
NO
Connection wires is correct?
YES
Please connect wires surely with an
external equipment according to the
paragraph 13-3.
NO
The address number is
correctly set.
YES
Please change RS-422/485
correctly.
NO
The communication
protocol is set correctly.
YES
Please make host's flow control
none.
NO
Command/data is set
correctly.
YES
Please set the address number
correctly referring to the paragraph
13-3.
NO
The flow control is specified
by the host.
YES
Please match the setting of
function correctly to the
specification of the host.
NO
The change of
RS-422/485 is set
correctly.
YES
Please use correct command/data
referring to the paragraph 13-3.
Inform Minebea about the contents of
failure and situation at site in details.
YES
NO
There may not respond because
host's response is slow. Please
enlarge the reply delay time of this
unit.
NO
The interval from the command
reception to the transmission of the
host is 1 ms or more.
The interval from command
reception to transmission of the
host is 1 ms or more.
Is setting at the reply
delay time appropriate to host's
response?
YES