DGA 9 7045899/02 EN 11.2020 Page 45
Recommendation for the mains connection cable:
Cable type: PUR-PUR CEE JB 3x1.50 (or similar)
Cable diameter : 8.1 mm
Cable cross-section : 1.5 mm
2
(of the 3 individual wires)
Minimum cross-section: 0.75 mm² (maximum cable length at 120V: 25m, at 230V: 50m
Modem communication:
(Exclusively for service purposes)
Ethernet communication:
Ethernet communication can be connected by copper cable (RJ45) or fibre optic cable (SC-Du-
plex) (please state when placing your order). For the copper variant, the connection socket (RJF
544) is located directly on the back of the device. The connection for the fibre optic version must
be routed through the cable gland and connected directly on the inside of the device.
Copper cable:
Recommendation for the Ethernet communication cable (copper):
Cable type: Ethernet cable Cat. 5e with RJ45 connector (RJF 544 6)
Cable diameter : 6.3 mm
Fibre optic cable:
Recommendation for the Ethernet communication cable (light waves):
Cable type: HITRONIC
®
HQN 1000 multi-mode G4 62.5/125 with SC-Duplex
connector
Cable diameter : 6.5 mm
Bending radius: min. 15 x Ø
Alarm outputs:
Recommendation for the analogue line cable:
Cable type: ÖLFLEX
®
FD855CP (or similar)
Cable diameter : 13.3 mm
Cable cross-section : 0.5 mm
2
(of the 16 individual wires)