MSENSE
®
DGA 9 7045899/02 EN 11.2020 Page 50
Note: In order to deactivate the MSENSE® DGA 9, the term “unmount” is typically used.
4.4.5 Automatic activation following the interruption of the supply voltage
If a device has already been activated, the automatic activation will be started after the supply voltage
is interrupted. You will hear a short beep sound and, after a brief moment, the screen will show the
special menu item “Extraction chamber (Pollution warning! - Indication of pollution!)”.
The waiting time until automatic activation is then started. This is shown with a second-by-second
countdown timer. Furthermore, short beeps are audible (the first 25 seconds is a short beep tone every
5 seconds and a short beep tone is then heard every second for the last 5 seconds).
There is the option to cancel the automatic activation with the key within 30 seconds and to
manually activate the device at a later point in time (4.4.3).
Once the waiting time has elapsed, the automatic activation will start and the LCD will show
the main menu.
Cancel the device’s automatic activation
Initial position of the cursor (the key has no stored function)
Exit menu
Remark: After leaving the initial position, this can no longer be reached.