GALAXY GX2
56
Using the GALAXY GX2
MSA
US
5.6 E-mail Notification
If a network connection is active and e-mail notification is co
nfigured via the MSA Link Pro
software, the Master Test Stand will send e-mails based on the events shown in the table below.
The MSA Link Pro user can configure which day(s) of the week that e-mails are sent and for which
conditions. Note that "EO" = Each Occurrence and "FO" = First Occurrence.
5.7 Overdue Calibration and Bump Notification
The Test Stand will determine that gas detectors are overdue for a bump or calibration and send
notification per the table above. Gas detectors are flagged as overdue on the day following the
due date.
Frequency Condition for Email e-mail is sent when
FO Low Calibration Gas Gas cylinder pressure 99 psi
FO Empty Calibration Gas Gas cylinder pressure 49 psi
FO Calibration Gas Nearing Expiration Date User configurable (default = 4 weeks)
FO Calibration Gas Expired Day of expiration
FO GX2 Faults/Errors As they occur
FO SD Memory Card Nearing Full 90% Full
FO SD Memory Card Full 99% Full
EO Security Accessed After password is entered
EO Bump Overdue User configurable (default = 1 day)
EO Calibration Overdue User configurable (default = 30 days)
EO Bump Failed As they occur
EO Calibration Failed As they occur
EO X-Cell Sensor Status Reported after 2 successive
failed calibrations

Table of Contents