5.6 E-mail Notification
If a network connection is active and e-mail notification is co
nfigured via the MSA Link Pro
software, the Master Test Stand will send e-mails based on the events shown in the table below.
The MSA Link Pro user can configure which day(s) of the week that e-mails are sent and for which
conditions. Note that "EO" = Each Occurrence and "FO" = First Occurrence.
5.7 Overdue Calibration and Bump Notification
The Test Stand will determine that gas detectors are overdue for a bump or calibration and send
notification per the table above. Gas detectors are flagged as overdue on the day following the
due date.
Frequency Condition for Email e-mail is sent when
FO Low Calibration Gas Gas cylinder pressure ≤99 psi
FO Empty Calibration Gas Gas cylinder pressure ≤49 psi
FO Calibration Gas Nearing Expiration Date User configurable (default = 4 weeks)
FO Calibration Gas Expired Day of expiration
FO GX2 Faults/Errors As they occur
FO SD Memory Card Nearing Full 90% Full
FO SD Memory Card Full 99% Full
EO Security Accessed After password is entered
EO Bump Overdue User configurable (default = 1 day)
EO Calibration Overdue User configurable (default = 30 days)
EO Bump Failed As they occur
EO Calibration Failed As they occur
EO X-Cell Sensor Status Reported after 2 successive
failed calibrations