161
EN
hyperion - TECHNICAL MANUAL
4.2.2 “HYB_CONN” codes - Ethernet connection errors
Cod. Meaning Description/solution
HYB_CONN 1 Message too long
Software version not compliant with mainboard rmware ver-
sion | check ethernet connection (cables / connectors / hub /
switch / routers / wi reachability)
HYB_CONN 2 Bad message size
Software version not compliant with mainboard rmware ver-
sion | check ethernet connection (cables / connectors / hub /
switch / routers / wi reachability)
HYB_CONN 3 Bad message format
Software version not compliant with mainboard rmware ver-
sion | check ethernet connection (cables / connectors / hub /
switch / routers / wi reachability)
HYB_CONN 4 Device has closed the connection Restart the device
HYB_CONN 5
Device does not responding to
message
Restart the device
HYB_CONN 6 Device not responding to ping
Restart the device | verify the PC subnet | use direct connec-
tion between PC and device
HYB_CONN 7
Connection timeout. Device not re-
sponding or not exist
Check device Ip address | verify the PC subnet | use direct
connection between PC and device
HYB_CONN 8
Socket Invalid. Send ethernet msg
without connection
Restart SW and check connection.
HYB_CONN 9
Ethernet not initialized. Send ether-
net msg without connection
Restart SW and check connection.
HYB_CONN 10 Wrong Device Model Check device Ip address | verify device model

Table of Contents