EasyManua.ls Logo

Myray hyperion X9 - PAN Sensor Calibration

Myray hyperion X9
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70
EN
hyperion - TECHNICAL MANUAL
2.3.4 PAN sensor calibration
OncetheX-Raybeamisnishedproceedwiththenextstep,thesensorcalibration.Clickon“PANsensorcalibration”
andpositionthealuminumlterinfrontofthecollimatorasshowninthepicturethatwillappear(seebelow).
Click on “start” and then on “Start procedure”. Press the X-Ray button every time the software prompts for it. The pro-
cedure is completely automatic.

Table of Contents

Related product manuals