hyperion - TECHNICAL MANUAL
4.2.4 “CALIB” codes - 2D calibration tool errors
Cod. Meaning Description/solution
CALIB 1 Unexpected null pointer restart the application and retry
CALIB 2 try to reinitialize thread already running restart the application and retry
CALIB 3 Routine source missing restart the application and retry
CALIB 4 Bad condition
Wait and retry | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 5 Not allowed request
Wait and retry | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 6 Dose out of limit
Wait and retry | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 7 Control not taken
Wait and retry | Take control |verify other device con-
nected | Disconnect and reconnect the device | restart
the application | Restart device
CALIB 8 COM port error
Wait and retry | verify COM port | Verify RS232 cable |
restart the application
CALIB 9 Wrong data from COM port
Wait and retry | verify COM port | Verify RS232 cable |
restart the application
CALIB 10 COM port time out
Wait and retry | verify COM port | Verify RS232 cable |
restart the application
CALIB 12 Leave control failed.
Wait and retry | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 13 Not allowed request
Verify connection | Wait and retry | Disconnect and re-
connect the device | restart the application | Restart de-
vice
CALIB 14 Error opening le
Verify le existing, access rights (read/write), available
memory, name le
CALIB 15 Previous request pending
Wait and retry | stop procedure |Disconnect and recon-
nect the device | restart the application | Restart device
CALIB 16 Not allowed request
Wait and retry | Verify connection | verify other device
connected|Disconnect and reconnect the device | restart
the application | Restart device
CALIB 17 Not allowed request
Wait and retry | Verify connection | verify other device
connected|Disconnect and reconnect the device | restart
the application | Restart device
CALIB 18 Control not taken
Take control | Wait and retry | Verify other device con-
nected | Disconnect and reconnect the device | restart
the application | Restart device
CALIB 19 Download Data failed
Wait and retry | Verify connection | Disconnect and re-
connect the device | Restart device
CALIB 20 Structure eld missing
Check FW version and application version | restart the
application | Restart device
CALIB 21 Global data changed but not saved Verify/Change data | Press save data button |
CALIB 22 Wrong value, conversion failed
Verify/Change data | reload previous value | Disconnect
and reconnect the device
CALIB 23 Refresh data still running
Wait and retry | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 24 Board missing
Check FW version and application version | Verify bo-
ards and cable connections
CALIB 25 Timeout request, previous request pending
Wait and retry | Stop procedure | Disconnect and recon-
nect the device | restart the application | Restart device