Cod. Meaning Description/solution
CALIB 26 Error writing le
Verify le existing, access rights (read/write), available
memory, name le
CALIB 27 Error reading le
Verify le existing, access rights (read/write), available
memory, name le
CALIB 28 Upload data failed
Wait and retry | Verify connection | Disconnect and re-
connect the device | Restart device
CALIB 29 Store data failed
Wait and retry | Verify connection | Disconnect and re-
connect the device | Restart device
CALIB 30 Read potenziometer failed
Wait and retry | Verify boards and cable connections |
Check potentiometer | Disconnect and reconnect the de-
vice | Restart device
CALIB 31 Wrong value
Wait and retry | Check potentiometer | Check cable |
Check board
CALIB 32 Read e2prom failed
Wait and retry | Take control | Disconnect and recon-
nect the device | restart the application | Restart device
| Check board
CALIB 33 Write e2prom failed
Wait and retry | Take control | Disconnect and recon-
nect the device | restart the application | Restart device
| Check board
CALIB 34 Dialog source missing restart the application and retry
CALIB 35 Remove le error
Verify le existing, access rights (read/write), available
memory, name le | Wait and retry | restart the applica-
tion
CALIB 36 Copy le error
Verify le existing, access rights (read/write), available
memory, name le | Wait and retry | restart the applica-
tion
CALIB 37 Beam width test failed
Wait and retry | press reset and retry | Check collimators
| restart the application | Restart device
CALIB 38 Wrong image received
remove undesidered objects between xray source and
detector and retry | Wait and retry | Download last image
and retry | Disconnect and reconnect the device | Verify
collimators | Restart device
CALIB 39 Wrong value in nasion calibration
remove undesidered objects between xray source and
detector and retry | Verify nasion position | Restart pro-
cedure | Verify collimators | Disconnect and reconnect
the device | Restart device
CALIB 42 Source code Case missing restart the application and retry
CALIB 43 Previous step is running
Wait and retry | Stop previous procedure | Disconnect
and reconnect the device | Restart device | restart the
application
CALIB 44 Wrong data
Verify/Change data | reload previous value | Disconnect
and reconnect the device
CALIB 45 Image totally out of center
Verify collimator positioning and restart procedure | re-
move undesidered objects between xray source and de-
tector and retry | restart device | restart the application
CALIB 46 Image not present
Wait and retry | Verify connection | Verify collimator po-
sitioning | remove undesidered objects between xray
source and detector and retry | Disconnect and recon-
nect the device | Restart device
CALIB 47 Timeout request
Wait and retry | Stop procedure | Disconnect and recon-
nect the device | restart the application | Restart device