Cod. Meaning Description/solution
CALIB 48 Wrong value, curtains limit passed
Verify curtains/collimator positioning and movement| Ve-
rify/Change data | reload previous value | Verify/change
curtains limit
CALIB 51 Wrong image type or size
Wait and retry | stop procedure | Download last image
and retry | Disconnect and reconnect the device | Restart
device
CALIB 52 Sensor afe gain out of tolerance
Verify alignment | Remove any object between tube and
sensor | Check sensor | Restart device
CALIB 53 Afe gain value not found
Verify alignment | Remove any object between tube and
sensor | Check sensor | Restart device
CALIB 54 Too many blemish lines found
Verify alignment | Remove any object between tube and
sensor | Check sensor | Restart device | Check limit va-
lue for blemish
CALIB 58 Makedir failed
Verify dir existing, access rights (read/write), available
memory, name dir
CALIB 61 Take control failed
Take control | Wait and retry | Verify other device con-
nected | Disconnect and reconnect the device | restart
the application | Restart device
CALIB 62 GetConguration failed
Wait and retry | Verify connection | Disconnect and re-
connect the device | restart the application | Restart de-
vice
CALIB 63 SetConguration failed
Wait and retry | Verify conguration data | Verify connec-
tion | Disconnect and reconnect the device | restart the
application | Restart device
CALIB 64 Wrong value pixel for mean (algo afe gain)
Verify data | Verify connection | Download last image and
retry |Check sensor |Restart device
CALIB 65 Wrong value graylevel
Verify data | Verify parameters | Verify connection | Di-
sconnect and reconnect the device | restart the applica-
tion | Restart device
CALIB 66 Wrong parameter
Verify parameters | Insert correct value and retry | reload
previous value | Disconnect and reconnect the device |
restart the application | Restart device
CALIB 67 Download image failed
Verify connection | Wait and retry | Disconnect and re-
connect the device | restart the application | Restart de-
vice
CALIB 68 SaveImage failed
Verify dir existing, access rights (read/write), available
memory, name le
CALIB 69 IP address missing Fill IP address eld
CALIB 70 Invalid IP address Insert valid IP address
CALIB 71 Upload sensor data failed
Wait and retry | Verify connection | Disconnect and re-
connect the device | Check sensor | Restart device
CALIB 72 Wrong parameter for sensor calibration
Verify data | Verify connection | Disconnect and recon-
nect the device | restart the application | Restart device
CALIB 73 Different version Update mainboard FW
CALIB 74 Different version Update the application
CALIB 75 Device is congured for a new language Set another language
CALIB 76 Restart device needed Restart device
CALIB 77 No device found
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 78 Find device failed
Wait and retry | Verify connection | Restart device | re-
start the application