hyperion - TECHNICAL MANUAL
Cod. Meaning Description/solution
CALIB 79 Wake up failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 80 Deserialize failed
Verify connection | Check version | restart the application
| Restart device
CALIB 81 Download sensor data failed
Wait and retry | Verify connection | check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 82 Store sensor data failed
Wait and retry | Verify connection | Check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 83 Serialize failed
Verify connection | Check version | restart the application
| Restart device
CALIB 84 Acquire trajectory image failed
Wait and retry | Verify connection | Check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 85 Acquire offset image ofailed
Wait and retry | Verify connection | Check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 86 Acquire static image failed
Wait and retry | Verify connection | Check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 87 Acquire TDI image failed
Wait and retry | Verify connection | Check sensor | Di-
sconnect and reconnect the device | Restart device
CALIB 88 allocation memory failed restart the application
CALIB 89 Serialize text failed
Verify connection | Check version | restart the application
| Restart device
CALIB 90 Deserialize text failed
Verify connection | Check version | restart the application
| Restart device
CALIB 91 Reset axes failed
Wait and retry | Verify connection | Check sensor | Re-
start device | restart the application
CALIB 92
Wrong value for image size in sensor calibra-
tion
Verify data | Verify connection | Download last image and
retry |Check sensor |Restart device | restart the applica-
tion
CALIB 93 Download Raw Image failed
Wait and retry | Verify connection | Check sensor | Re-
start device | restart the application
CALIB 94 Routine source missing restart the application
CALIB 95 XRayShot failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 96 Get data converter failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 97 Set data converter failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 98 Wrong converter le size
Verify data | Verify connection | Check cable |Restart de-
vice
CALIB 99 Set demo mode failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 100 Reset demo mode failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 101 Update tool failed
Wait and retry | Verify connection | Restart device | re-
start the application
CALIB 102 Move collimator failed
Wait and retry | Verify data | Verify connection | Verify
collimator | Reset axes | Restart device | restart the ap-
plication
CALIB 103 Send projection failed
Wait and retry | Verify connection | Restart device | re-
start the application