EasyManua.ls Logo

Newport e360 - Page 24

Newport e360
129 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 2-11 Software Download/Diagnostics Mode Screen
7
When software download is completed, ventilator will sound a
short beep. “
Download Complete
message will appear on
the screen and flashing stops.
8 Turn OFF the power of the e360 ventilator.
9 Wait at least 10 seconds and turn the e360 ventilator power
back ON.
10 Wait until “Ventilation Standby screen appears.
11 Press “Start V
entilating
to begin ventilating.
12 NOTE: Because Air/O
2
gas supplies are not connected you
will get an “Air/O
2
Loss” alarm. Disregard alarm while verifying
software is installed correctly.
13 After “Start Ventilation button is pressed, confirm that
Incompatible Software …. Message does not show up on
the screen. If this message appears on the LCD touch screen,
then the software download was NOT completed successfully.
Repeat 1 – 6 again. If the same problem still remains, then
contact NMI representative for further assistance.
14 Press “Extended Functions button on the Front control
panel.
15 Press “Event History button on LCD touch screen.
16 Confirm software is properly upgraded by checking software
version number (see Figure 2-12, left upper corner).
MAINTENANCE, OVERHAUL & SOFTWARE UPGRADE