Figure 2-12 Event History Log Screen
17
The new Software version is now downloaded successfully.
18 Remove USB Flash Drive from the USB Port on the back of
the e360 ventilator.
Circuit Check Test and Diagnostic
WARNING: Before returning the e360 ventilator back for patient
use after the software upgrade, you must perform the following:
1 Turn OFF the power of the e360 ventilator.
2 Connect medical grade air source to Air Inlet on the Back
Panel of the e360 ventilator and make sure that medical grade
air source can provide at least 30 psi and does not exceed
more than 90 psi. (see Figure 2-13 for
Air Inlet).
3 Connect appropriate patient circuit to the e360 ventilator as
shown in Figure 2-14. Occlude the patient circuit using a CAP.
A
filter and test lung are not required for this instruction.
4 Turn the power of the e360 ventilator back ON.
5 W
ait until ā€œ
V
entilation Standby
ā€ screen appears.
6 Follow the instructions that appear on LCD touch screen as
shown in Figure 2-15.
SECTION 2
SER360 A1106 2-13