22 Press “Manual Inflation” button to start Exhalation Flow
Sensor calibration,
d9. Wait for
30 sec ~ 1 minute
until “TSI
Air flow
:“ and “Exhale Flow:“ numbers start to increase
gradually and the air flow is delivered from the e360 ventilator.
It requires 6 ~ 10 minutes to complete Exhalation Flow Sensor
calibration, see Figure 2-18).
23 When flow delivered from the e360 ventilator stops and
“Message ID:” number changes to “41”, then the Exhalation
Flow Sensor calibration is successfully completed, see Figure
2-18.
24 Turn OFF the power of the e360 ventilator.
25 Disconnect medical grade air source from Air Inlet on the
Back Panel of the e360 ventilator.
26 Disconnect AC power cord from AC wall outlet.
27 Upon completion of the installation of software upgrade, you
must complete the attached “Newport e360 Ventilator
Software Update Form” with serial numbers, part numbers
and signatures and return it to Newport Regulatory
Department via Fax at +1.714.427. 0839.
If you have any questions please contact Newport Technical
Service Department:
Tel: +1.714.427.5811 ext. 500, Fax: +1.714.427-0572
Email: TechService@NewportNMI.com