EasyManua.ls Logo

Newport e360 - Page 76

Newport e360
129 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the supply is capable of supporting a minimum of 180 L/min of
constant flow.
Attach the patient circuit to the e360 ventilator.
NOTE: Observe ESD precautions throughout the following
procedures for Analog PCB and Pneumatics Calibrations:
Analog PCB Calibration: d5
To locate the transducers on the Analog PCB, see Figure 5-2, below.
1 d0 will be displayed in the “Trig window on the Control
Panel and “0 will be displayed in the “Flow/t Insp window
when the e360 ventilator enters into the Diagnostics Mode.
2 Press the button below the “Trig window, then the display
will start blinking.
3 While the display is blinking, turn the Adjustment Knob to
change the setting from “
d0 to “d5”.
Then press the
Accept button to acknowledge.
NOTE: Steps 4 to 14 are to calibrate zero pressure offset of
the Exhalation drive pressure transducer
XD107.
4 Remove the ORANGE tubing from transducer XD107 on the
Analog PCB.
5 Press the “Manual Inflation button to start the of
fset
calibration. The Message ID will change to “25”. Refer to
Table 5-5 for the descriptions of all message ID’s.
6 Verify the setting of “t Insp is “0”. If not, press the button below
Flow / t Insp window, then the display will start blinking.
7 While the display is blinking, turn the Adjustment Knob to
adjust the setting of “t Insp to “0”, and then press the
Accept button to acknowledge.
8 Press the button below “Resp Rate window
, then the display
will start blinking.
9 While the display is blinking, turn the Adjustment Knob to
change the setting to any value, and then press the “
Accept
button to acknowledge.
10 After the “Accept button is pressed, the value of “Drive
Pressure displayed on the LCD touch Screen will change.
CALIBRATION PROCEDURES
5-4 SER360 A1106

Table of Contents

Other manuals for Newport e360

Related product manuals