11 As the setting of “Resp Rate increases, the value of “Drive
Pressure
will decrease after the “Accept button is pressed.
12 As the setting of “Resp Rate decreases, the value of “Drive
Pressure will increase after the “Accept button is pressed.
13 When the setting of “Resp Rate is “0”, which cannot be
further decreased, consequently the value of “Drive Pressure”
cannot be increased. In this case, first increase the setting of
Resp Rate to a high value and then decrease it. Always
press “Accept” button when the adjustment of the “
Resp Rate
setting is made.
14 Repeat steps 8 to 13 to obtain drive pressure of 3.00 cmH
2
O
±
0.05cmH
2
O, then press “Manual Inflation button to
acknowledge. The Message ID will change to “26”.
NOTE: Steps 15 to 25 are to calibrate the gain of the
Exhalation drive pressure transducer
XD107.
15 Press the button below “Flow / t Insp window, then the
display will start blinking.
16 While the display is blinking, turn the Adjustment Knob to set
the value of “
t Insp to “1”, and then press the “Accept
button to acknowledge.
17 Use a pressure syringe to apply a pressure of 110 ± 1 cmH
2
O
to the bottom port of transducer XD107. (T
ee in a pressure
gauge 0 to 200 cmH
2
O to insure accurate pressure delivery).
18 Press the “Manual Inflation button to start the gain
calibration.
The Message ID will change to “25”.
19 Press the button below the “Resp Rate window, then the
display will start blinking.
20 While the display is blinking, turn the Adjustment Knob to change
the setting to any value, then press the “
Accept button.
21 After the “Accept button is pressed, the value of “Drive
Pressure
displayed on the LCD touch Screen will change.
22 As the setting of “Resp Rate increases, the value of “Drive
Pressure will
increase after the “
Accept button is pressed.
23 As the setting of “Resp Rate
decreases
, the value of “Drive
Pressure
will decrease after the “Accept” button is pressed.
24 Repeat steps 19 to 23 to obtain drive pressure of 1
10.0
cmH
2
O ± 1.0 cmH
2
O, then press “Manual Inflation button
to acknowledge. The Message ID will change to “26”.