11 As the setting of “Resp Rate” increases, the value of “Drive
Pressure
” will decrease after the “Accept” button is pressed.
12 As the setting of “Resp Rate” decreases, the value of “Drive
Pressure” will increase after the “Accept” button is pressed.
13 When the setting of “Resp Rate” is “0”, which cannot be
further decreased, consequently the value of “Drive Pressure”
cannot be increased. In this case, first increase the setting of
“
Resp Rate” to a high value and then decrease it. Always
press “Accept” button when the adjustment of the “
Resp Rate”
setting is made.
14 Repeat steps 8 to 13 to obtain drive pressure of 3.00 cmH
2
O
±
0.05cmH
2
O, then press “Manual Inflation” button to
acknowledge. The Message ID will change to “26”.
NOTE: Steps 15 to 25 are to calibrate the gain of the
Exhalation drive pressure transducer
XD107.
15 Press the button below “Flow / t Insp” window, then the
display will start blinking.
16 While the display is blinking, turn the Adjustment Knob to set
the value of “
t Insp” to “1”, and then press the “Accept”
button to acknowledge.
17 Use a pressure syringe to apply a pressure of 110 ± 1 cmH
2
O
to the bottom port of transducer XD107. (T
ee in a pressure
gauge 0 to 200 cmH
2
O to insure accurate pressure delivery).
18 Press the “Manual Inflation” button to start the gain
calibration.
The Message ID will change to “25”.
19 Press the button below the “Resp Rate” window, then the
display will start blinking.
20 While the display is blinking, turn the Adjustment Knob to change
the setting to any value, then press the “
Accept” button.
21 After the “Accept” button is pressed, the value of “Drive
Pressure
” displayed on the LCD touch Screen will change.
22 As the setting of “Resp Rate” increases, the value of “Drive
Pressure” will
increase after the “
Accept” button is pressed.
23 As the setting of “Resp Rate”
decreases
, the value of “Drive
Pressure
” will decrease after the “Accept” button is pressed.
24 Repeat steps 19 to 23 to obtain drive pressure of 1
10.0
cmH
2
O ± 1.0 cmH
2
O, then press “Manual Inflation” button
to acknowledge. The Message ID will change to “26”.
SECTION 5
SER360 A1106 5-5