Note: If pressure is only applied to transducer XD105 during this
calibration, Device Alert will occur when the pressure to XD105 is
20cmH2O higher than the pressure to XD106. Device Alert can be
only cleared by turning OFF and ON the e360 ventilator.
30 To calibrate the zero pressure offset of Air Gas Supply
pressure transducer XD102, repeat steps 4 to 14, except
for the following difference.
a. In step 4, remove
YELLOW tubing from transducer
XD102.
b. In step 7, adjust the setting of “
t Insp to “6”.
c. In step 14, calibrate the value of “
Air Gas Supply to
0.0
+
0.1 psig.
31 To calibrate the gain of Air Gas Supply pressure transducer
XD102, repeat steps 15 to 25, except for the following
difference.
a. In step 16, adjust the setting of “
t Insp to “7”.
b. In step 17, apply a pressure of 25psig to the
top
port
of transducer
XD102.
c. In step 24, calibrate the value of “
Air Gas Supply to
25.0 ± 0.1 psig.
d. In step 25, connect
YELLOW tubing back to the top
port of transducer XD102.
32 To calibrate the zero pressure offset of Oxygen Gas Supply
pressure transducer
XD103, repeat steps 4 to 14, except for
the following difference.
a. In step 4, remove
BLUE tubing from transducer XD103.
b. In step 7, adjust the setting of “t Insp to “8”.
c. In step 14, calibrate the value of “
Oxygen Gas
Supply
to 0.0
+
0.1 psig.
33 T
o calibrate the gain of Oxygen Gas Supply pressure
transducer
XD103, repeat steps 15 to 25, except for the
following difference.
a. In step 16, adjust the setting of “
t Insp to “9”.
b. In step 17, apply a pressure of 25psig to the top
port
of transducer
XD101.
SECTION 5
SER360 A1106 5-7