c. In step 24, calibrate the value of “Oxygen Gas
Supply
to 25.0 ± 0.1 psig.
d. In step 25, connect
BLUE tubing back to the top port
of transducer
XD103.
34 Turn OFF the power, and then disconnect the Air and O
2
gas
supplies and AC power.
Figure 5-2 Analog PCB
CALIBRATION PROCEDURES
5-8 SER360 A1106