7.1 Specification
7-3 ZM2376
Measurement conditions
Measurement frequency 1mHz to 5.5MHz, Resolution 6 digits (1mHz when < 100Hz),
Accuracy ±0.01%
Measurement signal
Level Setting range 10mV to 5V, Resolution 3 digits (1mV when < 100mV)
Level accuracy ± (8% + 5mV) 1MHz (typical value when < 1Hz)
± (10% + 5mV) > 1MHz
RMS values at open output.
ALC Constant voltage drive / Constant current drive / Disabled
Voltage setting range 10mV to 5V, Resolution 3 digits (1mV when <
100mV)
Current setting range 1µA to 200mA, Resolution 3 digits (0.1µA when <
10µA)
The constant control range will be narrower than the above specifications
depending on the product dispersion or DUT’s impedance.
The current range is restricted depending on the measurement range.
Output impedance 6 / 25 / 100 (supplementary value).
The minimum output impedance can be selected from the above three values.
However, it is limited by the measurement range, the signal level, and the
frequency.
Internal DC bias Setting range 0V to +5V, Resolution 1mV,
Accuracy ± (2% + 5mV) [Level 2V]
± (2% + 10mV) [Level 2V]
at signal frequency of 1 MHz or less, 23±10°C (Environmental
temperature), and open output.
It can be turned on / off.
Trigger source INT Internal (automatic continuous trigger)
MAN Manual
EXT Handler interface
BUS Remote control
Trigger delay time Setting range 0s to 999.9999s, Resolution 0.0001s
(Time after input of trigger until start of signal acquisition)
Triggered drive Drive only at measurement / Continuous drive selectable
(Measurement signal can be output only during the time from trigger to
completion of signal acquisition)