104
Chapter 4 Reprocessing Operations
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5. Connect the leak test air tube connector (black) to the reprocessing basin’s
leak test connector (black).
Figure 4.26
6. Close the lid by pushing it until it clicks.
Including leak test in a reprocessing program
An endoscope leak test can be incorporated in a reprocessing program.
1. Perform Step 1 of the procedure in Section 4.9, “Reprocessing” on
page 113.
2. Make sure that the water faucet is open, and then press the LEAK TEST
button on the main control panel to select LEAK TEST, and press the
START button. The water supply will start and water will fill the reprocessing
basin.
Figure 4.27
190 series endoscope
1
180 or older series endoscope
1 190 series endoscopes may not be available in some areas.