EasyManua.ls Logo

Olympus OER-Pro - Replacing the Detergent Tank

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Inspection Before Use
45
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Replacing the detergent tank
Before handling the detergent, read the precautions and
instructions for use carefully, and use the detergent as
instructed. Make sure that you fully understand what
measures need to be taken if you get any detergent on your
skin.
Always use an Olympus-validated detergent. Otherwise, the
endoscope may not be properly cleaned and as a result, the
endoscope may not achieve high-level disinfection.
When handling the detergent, always wear appropriate
personal protective equipment, such as eyewear, face mask,
moisture-resistant clothing, and chemical-resistant gloves
that fit properly and are long enough so that your skin is not
exposed. All personal protective equipment should be
inspected before use and replaced periodically before it is
damaged.
To prevent the detergent from leaking, do not tilt the
detergent tank when there is detergent inside.
Do not add detergent into the detergent tank. Detergent tank
is disposable and it is not intended to be reused.
If reprocessing is initiated without detergent, the error code
[E95] is displayed and the reprocessing is stopped. The
detergent replacement indicator on the main panel blinks.
Contact Olympus to obtain the list of Olympus-validated
detergents.
The detergent tank can hold about 2 L (67 ounces) of
detergent (which can be used for about 30 reprocessing
operations).
When reprocessing is performed with newly-installed
equipment or with equipment after Section 7.16, “Preparing
the reprocessor for long-term storage” on page 259 is
performed, reprocessing may stop with the error code [E95]
displayed even though there is enough detergent in the tank.
See page 125 (“If the error code [E95] is displayed again
after replacing the detergent tank and restarting the
reprocessing process:”) to solve this problem.