Appendix
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Shipping environment
Operating environment
Purified water is water that has been produced through the
methods of either reverse osmosis (RO), deionization (DI),
distillation or other methods that meet USP standards to
remove impurities.
Shipping
environment
Ambient temperatures –47 to +60C (–52 to +140F)
Relative humidity 10 – 90
Atmospheric pressure 700 – 1060 hPa
Operating
environment
Ambient temperatures 10 – 40C (50 – 104F)
Relative humidity 30 – 85
Elevation 3000 meters (maximum)
Designed for use Indoors
Water supply flow 17 l/min. or more when the water faucet is
fully open
Water supply pressure 0.1 – 0.5 MPa (include water hammer)
Water supply
temperature
Max. 28C
Water type Potable water or Softened water
(Purified water cannot be used)
Water hardness 400 ppm (maximum)
0 – 150 ppm (Recommended value)
1
1 The recommended value is in reference to AAMI TIR34:2007 “Water for
the reprocessing of medical devices”.