Chapter 5 End-of-Day Checks
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
Chapter 5 End-of-Day Checks
To ensure safe, reliable operation, inspect and clean all parts of the device
regularly.
Be sure to inspect and clean the equipment as described in
this chapter. Otherwise, the functions and performance of the
equipment may not operate properly.
If any irregularity is observed, do not use the device and
contact Olympus. If the device is used when an irregularity is
observed, the device may malfunction. Water leakage,
electric shock, burns, and/or fire may also result.
When inspecting the device, always wear appropriate
personal protective equipment, such as eyewear, face mask,
moisture-resistant clothing, and chemical-resistant gloves
that fit properly and are long enough so that your skin is not
exposed. All personal protective equipment should be
inspected before use and replaced periodically before it is
damaged.
Check Checks at the end of every working day
5.1 Turning the power OFF, closing the water faucet and cleaning the outer
surface
5.2 Cleaning the mesh filters
5.3 Cleaning the fluid level sensor
Table 5.1
Check Required items
70% ethyl alcohol or isopropyl alcohol
Clean cloth
Filter cleaning brush
Table 5.2