Chapter 7 Routine Maintenance
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
9. When air purge has started, touch the air filter connectors to ensure that air
is not leaking out. Also, ensure that the connectors do not produce a
whistling sound, which would mean there is an air leak.
Figure 7.48
10. Press the STOP/RESET button to end air purge. If an air leak is detected,
reinstall the air filter as described in Section 3.10, “Installation of air filter
(MAJ-823)” in the “Instructions-Installation Manual”.
11. Close the front door.
7.6 Cleaning the float switch
Always press the power switch OFF before cleaning the float
switch. Moving the float switch while the power switch is ON
will be detected as an error by the device and result in error
processing.
1. Make sure that the power switch is OFF.
Check Required items
Detergent
Brush
Table 7.10