Chapter 5 End-of-Day Checks
149
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
5.1 Turning the power OFF, closing the water faucet
and cleaning the outer surface
To prevent water leakage, be sure to close the water faucet
at the end of the working day.
After using the device, dry it thoroughly (so that no water
remains in the reprocessing basin) and close the lid before
storage. Otherwise, microorganisms may proliferate in the
device.
If the equipment has been stored after closing the lid without
drying the reprocessing basin completely, thoroughly wipe
the inside of the reprocessing basin with a cloth moistened
with 70% ethyl alcohol or isopropyl alcohol before the next
use.
1. Close the water faucet.
2. Press the power switch to OFF.
3. Using a clean cloth moistened with neutral detergent, clean every part of the
device including the front and back of the lid, the lid packing, the edge and
inside of the reprocessing basin and the control panel, then wipe with a dry
clean cloth. To prevent growth of microorganisms, it is also recommended
to wipe every part of the equipment with a cloth moistened with 70% ethyl
alcohol or isopropyl alcohol.
Figure 5.1