EasyManua.ls Logo

Olympus OER-Pro - 6.10 Emergency Stop and Auto Processing

Olympus OER-Pro
308 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
172
Chapter 6 Other Functions
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
6.9 Alcohol flushing
This process automates the flushing of the endoscope channels with alcohol
followed by air to help dry the channels.
When using the disinfectant solution and alcohol, Olympus
recommends the use of gas filters and running this
equipment in well-ventilated areas.
Wear a face mask, gloves, and protective clothes to
minimize aspiration and skin contact.
Wear goggles for eye protection.
When the alcohol flushing process is stopped due to an
equipment error, do not use the endoscope and start the
alcohol flushing process again from the beginning.
Otherwise, alcohol may remain in the endoscope channel
and pose a risk to patient safety.
Do not perform alcohol flushing without connecting the
connecting tubes. Otherwise, excessive pressure on the
pipes in the equipment may damage it.
Refer to the following association’s guidelines related to
ventilation:
SGNA (Society of Gastroenterology Nurses and
Associates)
ASGE (American Society of Gastroenterological
Endoscopy)
APIC (Association for Professionals of Infection
Control and Epidemiology)
AORN (Association of Preoperative Registered
Nurses)
ASTM (American Society for Testing and Materials)
OSHA (Occupational Safety and Health
Administration)
ACGIH (American Conference of Governmental
Industrial Hygienists)
NIOSH (National Institute for Occupational Safety
and Health)
AIA (American Institute of Architects)
Do not handle the equipment if the operator shows any
allergic symptoms even while wearing protective gear.