Chapter 7 Routine Maintenance
259
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
7.16 Preparing the reprocessor for long-term storage
When the equipment will be stored for more than 14 days, follow the procedure
described in this section.
Before handling the alcohol, read the precautions carefully
and use it as instructed.
Do not tilt the alcohol tank while there is still alcohol inside.
Otherwise, the alcohol may spill.
1. Discharge disinfectant solution from the equipment as described in
“Draining through the disinfectant collection hose” on page 226.
2. Pull out the detergent/alcohol drawer.
3. Detach the detergent tank cap to which a tube is connected (Do not
disconnect the connector.) (see Figure 7.98).
Figure 7.98
4. Step on the foot pedal to open the lid.
Check Required items
Syringe and tube
Connector jigs
70% ethyl alcohol or isopropyl alcohol
Container (small) with a capacity of about 200 ml, such as a beaker
Provided wrench
Clean cloth
Sterile water: more than 50 ml
Table 7.19