Chapter 7 Routine Maintenance
261
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
9. Put the cap in the container filled with the sterile water (see Figure 7.101).
Figure 7.101
10. To rinse the line, connect the tube attached to the syringe to the detergent
nozzle and suction more than 50 ml of sterile water with the syringe (see
Figure 7.100).
11. To remove rinse water from inside the line, connect the tube attached to the
syringe to the detergent nozzle inside the reprocessing basin. Suction the
water with the syringe until no more water comes out of the line (see Figure
7.100).
12. Prepare a beaker or similar container and pour about 50 ml of the 70% ethyl
alcohol or isopropyl alcohol into it.
13. Put the detached cap in the container of the 70% ethyl alcohol or isopropyl
alcohol (see Figure 7.101).
14. To rinse the line, connect the tube attached to the syringe to the detergent
nozzle and suction more than 50 ml of 70% ethyl alcohol or isopropyl
alcohol with the syringe (see Figure 7.100).
15. To remove 70% ethyl alcohol or isopropyl alcohol from inside the line,
connect the tube attached to the syringe to the detergent nozzle inside the
reprocessing basin. Suction the alcohol with the syringe until no more
alcohol comes out of the line (see Figure 7.100).