Chapter 7 Routine Maintenance
221
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
8. Turn the connectors to correct the orientations of the tubes as shown below.
Figure 7.52
9. Close the detergent/alcohol drawer.
7.8 Cleaning the accessories and accessory holder
As the accessories listed below tend to attract dirt and dust, they should be
cleaned periodically and stored in a clean environment. The accessory holder
used for storage should also be kept clean in the same way.
Connecting tubes
Filter tubes
Hoses
Gas filter case excluding the gas filter
Gas filter adapter, etc.
Accessory holder
Card holders
Check Required items
70% ethyl alcohol or isopropyl alcohol
Clean cloth
Table 7.11
Connectors

Table of Contents