Chapter 7 Routine Maintenance
251
ENDOSCOPE REPROCESSOR OER-Pro Operation Manual
7.15 Disinfecting the internal detergent/alcohol lines
When handling the disinfectant solution and
detergent/alcohol, carefully read the precautions for its use to
fully understand the given information and use as instructed.
Particular understanding is required for measures to be taken
in case the disinfectant solution comes into contact with your
skin.
When handling the disinfectant solution and
detergent/alcohol, wear appropriate personal protective
equipment to avoid direct contact with your skin or excessive
inhalation of its vapor. The disinfectant solution and its vapor
may affect the human body. Personal protective equipment,
such as eyewear, face mask, moisture-resistant clothing, and
chemical-resistant gloves that fit properly and are long
enough so that your skin is not exposed. All personal
protective equipment should be inspected before use and
replaced periodically before it is damaged.
To prevent spills, keep the detergent/alcohol tanks upright.
Draining and using the disinfectant solution from the
equipment
Collect the disinfectant solution as described in Section 3.10, “Checking the
disinfectant solution concentration level” on page 59. Always check the
disinfectant solution concentration and use it to disinfect detergent/alcohol
supply pipe only when it’s effective.
Disinfecting of the detergent supply line
Check Required items
Syringe and tube
Two Cup-shape container with a capacity of about 200 ml, such as a beaker
Disinfectant solution: approximately 30 ml
Sterile water: more than 100 ml
Table 7.17